Generic selectors
Exact matches only
Search in title
Search in content
Post Type Selectors
Filter by Categories
Acknowledgements
Authors’ reply
Book Review
Book Reviews
Classics In Indian Medicine
Clinical Case Report
Clinical Case Reports
Clinical Research Methods
Clinico-pathological Conference
Clinicopathological Conference
Conferences
Correspondence
Corrigendum
Editorial
Eminent Indians in Medicine
Errata
Erratum
Everyday Practice
Film Review
History of Medicine
HOW TO DO IT
Images In Medicine
Indian Medical Institutions
Letter from Bristol
Letter from Chennai
Letter From Ganiyari
Letter from Glasgow
Letter from London
Letter from Mangalore
Letter From Mumbai
Letter From Nepal
Masala
Medical Education
Medical Ethics
Medicine and Society
News From Here And There
Notice of Retraction
Notices
Obituaries
Obituary
Original Article
Original Articles
Review Article
Selected Summaries
Selected Summary
Short Report
Short Reports
Speaking for Myself
Speaking for Ourselve
Speaking for Ourselves
Students@nmji
Generic selectors
Exact matches only
Search in title
Search in content
Post Type Selectors
Filter by Categories
Acknowledgements
Authors’ reply
Book Review
Book Reviews
Classics In Indian Medicine
Clinical Case Report
Clinical Case Reports
Clinical Research Methods
Clinico-pathological Conference
Clinicopathological Conference
Conferences
Correspondence
Corrigendum
Editorial
Eminent Indians in Medicine
Errata
Erratum
Everyday Practice
Film Review
History of Medicine
HOW TO DO IT
Images In Medicine
Indian Medical Institutions
Letter from Bristol
Letter from Chennai
Letter From Ganiyari
Letter from Glasgow
Letter from London
Letter from Mangalore
Letter From Mumbai
Letter From Nepal
Masala
Medical Education
Medical Ethics
Medicine and Society
News From Here And There
Notice of Retraction
Notices
Obituaries
Obituary
Original Article
Original Articles
Review Article
Selected Summaries
Selected Summary
Short Report
Short Reports
Speaking for Myself
Speaking for Ourselve
Speaking for Ourselves
Students@nmji
View/Download PDF

Translate this page into:

Clinical Case Report
2021:34:2;84-85
doi: 10.4103/0970-258X.326761

Management of a complicated ‘pacemaker pocket’ site infection

Karthik Raghuram, Krishna Kumar Mohanan Nair, Narayanan Namboodiri, Debasish Das, Ajitkumar Valaparambil
 Department of Cardiology, Sree Chitra Tirunal Institute for Medical Sciences and Technology, Thiruvananthapuram 695011, Kerala, India

Corresponding Author:
Krishna Kumar Mohanan Nair
Department of Cardiology, Sree Chitra Tirunal Institute for Medical Sciences and Technology, Thiruvananthapuram 695011, Kerala
India
kakkam@gmail.com
Published: 28-Sep-2021
How to cite this article:
Raghuram K, Mohanan Nair KK, Namboodiri N, Das D, Valaparambil A. Management of a complicated ‘pacemaker pocket’ site infection. Natl Med J India 2021;34:84-85
Copyright: (C)2021 The National Medical Journal of India

Abstract

Pocket site infection after implantation of a pacemaker is a dreaded complication which requires removing the device and reimplanting it at a contralateral site. Difficulties arise when the patient is dependent on pacing and when there are issues with venous access at the contralateral site. We report a patient with pacemaker pocket site infection with congenital complete heart block managed with explantation of the device, semi-permanent pacing during antibiotic treatment, reimplantation of the device at the contralateral site and management of subtotal subclavian vein stenosis noted during reimplantation.

Introduction

Pocket site infection of a cardiovascular implantable electronic device (CIED) is treated by removal of the device. This is followed by complete treatment of the infection and reimplantation of the device in a new site. Management is complicated when the patient is dependent on pacing, requiring temporary pacing during the treatment of the infection and when there is difficulty in venous access at the new site.

The Case

A 16-year-old boy, known to have congenital complete heart block, presented with syncope in 2014 at a nearby hospital. A permanent pacemaker (VVI) was implanted through the right subclavian vein. The patient presented in June 2018, with complaints of syncope, was evaluated and found to have a lead fracture. An attempt at lead revision failed, the lead was abandoned on the right side, and a pacemaker (VVI) was placed through the left subclavian vein. He again presented in August 2018 to the same hospital with pain, swelling and pus discharge from the left-sided system pocket site. The patient was diagnosed to have a pacemaker pocket site infection and managed conservatively with antibiotics. In January 2019, the patient presented with recurrent symptoms of pain, swelling and pus discharge from the left-side pocket site. There was dehiscence of the pocket site. Initially he was managed with antibiotics, secondary suturing and referred to our hospital.

At presentation, the patient was afebrile with stable vital parameter. Localized pus discharge was seen at the incision site with sutures in situ. There was no erythema or induration. A trans-oesophageal echocardiography was done to exclude vegetations over the leads and valves. The chest X-ray showed the abandoned lead on the right side and the left-sided VVI system [Figure - 1]. Investigations done showed no leucocytosis, with normal C-reactive protein and procalcitonin levels. The pus and blood cultures were sterile. The patient was diagnosed to have an isolated generator pocket infection. The left-sided generator and lead were explanted. As the patient was pacing-dependent, a semi-permanent pacing strategy was implemented to provide pacing support during treatment of the infection. A right internal jugular vein access was used to place a screw-in lead in the right ventricle, and it was connected to an external generator [Figure - 2]. The patient was given intravenous antibiotics for 2 weeks and a new dual chamber pacemaker (DDD) was planned on the right side. At the time of venous puncture, subtotal right subclavian occlusion was present [Figure - 3]. Initially, a Teflon wire was used to cross the stenosis but was not successful. A 0.0322 Terumo wire was used to negotiate the stenosis. Over the Terumo wire, dilator of 6F sheath was advanced. A balloon dilatation was not done as the stenosis was negotiated using the Terumo wire and dilator. This was followed by the introduction of a 6F peel-away sheath. A 6F lead was then advanced through the sheath, lead was screwed in and the DDD system placed. The previously abandoned lead was left in situ [Figure - 4]. The patient received 2 weeks of antibiotics after the right-sided implant. Both sites healed well on follow-up. The abandoned lead is in situ and requires regular monitoring.

Figure 1: Chest X-ray showing abandoned lead on the right side and the left-sided VVI system
Figure 2: Semi-permanent pacing with right internal jugular vein access and screw-in lead in the right ventricle, with external pulse generator
Figure 3: Venogram showing subtotal occlusion of the right subclavian vein
Figure 4: Right-sided dual chamber pacemaker with previously abandoned lead in situ

Discussion

Pocket site infections occur at the rate of 4.82 per 1000 permanent pacemaker (PM)-years.[1] The reported admission mortality rates for all cardiovascular implantable electronic device (CIED) infections vary from 3.7% to 8.1%.[2] According to the 2017 Heart Rhythm Society expert consensus statement of CIED lead management and extraction, if there is a pocket infection, the CIED must be removed and antibiotics given for 2 weeks after removal. Reimplantation of the CIED can be done at a later date, depending on the clinical situation.[3]

With pacing-dependent patients, temporary pacing is required prior to reimplanting a new permanent device. The ‘semi-permanent’ pacing method includes using a screw-in pacing lead connected to an external pulse generator. This method ensures adequate pacing while the patient waits for re-implantation of a new device.[3]

New device implantation is usually recommended 72 hours to 14 days after explantation, depending on the clinical situation. If blood cultures are positive prior to explantation, the new device is delayed until the blood cultures are sterile for 72 hours. Reimplantation can be delayed further if there is a persisting clinically suspected source of infection. Reimplantation is done at a site contralateral to the initial device.[3]

Subclavian vein stenosis after pacemaker/implantable defibrillator has been reported with a variable incidence in the literature. Most studies have noted that a moderate degree of subclavian vein stenosis (>50% narrowing) occurs in 50% of patients. Even severe stenosis (>70%) can also be asymptomatic due to the development of venous collaterals. Difficulties arise when there is a need for lead revision. Most subtotal stenosis can be negotiated with a stiff guidewire. In more severe symptomatic cases, there may be a need for percutaneous venoplasty (with or without stenting) or surgical bypass.[4]

Conclusion

We have outlined the management of pacemaker pocket site infection in a patient who was dependent on pacing. We managed the subtotal subclavian vein stenosis that can be encountered at the contralateral site, especially in a subset of patients who had a previous intervention at the contralateral site. The operator should expect and be prepared for these complications.

Conflicts of interest. None declared

References
1.
Johansen JB, Jørgensen OD, Møller M, Arnsbo P, Mortensen PT, Nielsen JC, et al. Infection after pacemaker implantation: Infection rates and risk factors associated with infection in a population-based cohort study of 46299 consecutive patients. Eur Heart J 2011;32:991–8.
[Google Scholar]
2.
Sohail MR, Henrikson CA, Braid-Forbes MJ, Forbes KF, Lerner DJ. Mortality and cost associated with cardiovascular implantable electronic device infections. Arch Intern Med 2011;171:1821–8.
[Google Scholar]
3.
Kusumoto FM, Schoenfeld MH, Wilkoff BL, Berul CI, Birgersdotter-Green UM, Carrillo R, et al.2017 HRS expert consensus statement on cardiovascular implantable electronic device lead management and extraction. Heart Rhythm 2017;14: e503–51.
[Google Scholar]
4.
O’Leary B, Allaqaband S. Subclavian vein stenosis/occlusion following transvenous cardiac pacemaker and defibrillator Implantation: Incidence, pathophysiology and current management. J Patient Cent Res Rev 2015;2: 112–17.
[Google Scholar]